结合夏县中心地震台CIC-200型离子色谱仪实际观测,从仪器输液系统、分离系统、基线、管路等方面详细阐述色谱仪常见故障,逐一分析原因,并给出相应故障排除方法,为同类仪器维护提供参考,并为获得准确观测数据,延长仪器使用寿命,提出日常维护方法及建议。
Taking CIC-200 ion chromatograph as an example, combined with the actual observation of Xiaxian Central Seismic Station, the common failures of ion chromatograph in daily observation are described in detail from the aspects of instrument infusion system, separation system, baseline, pipeline and other problems, and the reasons are analyzed one by one. The corresponding elimination methods were put forward to provide reference for the usage of similar instruments.
2018,39(4): 164-168 收稿日期:2017-06-04
DOI:10.3969/j.issn.1003-3246.2018.04.022
基金项目:夏县中心地震台背景场新型模拟色谱观测系统分析及运维保障技术研究(项目编号:SBK-1732);离子色谱仪地震观测常见问题及解决方法(项目编号:CEA-JC/3JH-170504)
作者简介:常姣(1988-),女,工程师,主要从事地下流体监测预报工作。E-mail:495066985@qq.com