为了避免地震计由强振动引起的不同程度的位移、偏转或倾倒,研制出1套地震计固定装置,该装置使地震计能够经受强烈振动。与其他类似装置不同的是,该装置是通过地震计的3个底脚进行固定的,不与地震计其他位置接触,这有效避免了混入其他噪声。通过测试试验发现,在安装固定底座后,地震计运行和产出的波形均正常,噪声略有变小,平均信噪比变化不大。
In order to avoid the displacement, deflection, or toppling of a seismometer caused by strong vibration, a set of seismometer fixing devices was developed to ensure that the seismometer can withstand strong vibration. Different from other similar devices, the seismometer is fixed by the device with its three feet and does not contact other parts of the seismometer, which effectively avoids introducing other noises. After installing the fixing devices, the seismometer works normally, and the waveform of the seismometer is normal. The noise becomes slightly smaller and the average signal-to-noise ratio does not change much.
2021,42(1): 132-138 收稿日期:2020-07-13
DOI:10.3969/j.issn.1003-3246.2021.01.019
基金项目:河北省地震科技星火计划面上项目(项目编号:DZ20190418019)
作者简介:张磊(1981—),男,高级工程师,主要从事地震监测和仪器运维工作。E-mail:zhanglei14324@163.com
参考文献:
吕俊强,杜文勇,许文俊. 宁夏测震台网的大震应对改进方案[J]. 华南地震,2010,30(4):55-59.
肖武军,张尧,王立森,等. 地震台站标准化设计在冬奥会保障项目中的应用——以河北阳原台为例[J].中国地震,2019,35(3):565-572.
张新东,闫俊岗,张小涛. 地震计的固定[J]. 地震地磁观测与研究,2013,34(3/4):186-187.
中国地震局.中国数字测震台网技术规范[M].北京:地震出版社,2005.
中国地震局监测预报司.地震学与地震观测[M].北京:地震出版社,2007.
中国地震局监测预报司.关于开展2019年地震台站标准化改造工作的通知[Z]. 2019